Each channel has a designated enable/disable button. When individual
channels are active or enabled the buttons are illuminated. This allows
for easy reference and avoids any confusion of output operability. The
run/stop indicator on the front panel LCD display as well as an illuminated
run/stop button further simplify setup.
The Model 575 offers additional inputs and outputs for external clock
syncing. Specify your input / output reference frequency (10MHz to
100MHz). Sync with the Mode Lock Oscillator of a laser, or phase lock
multiple units with one clock.
Illuminated Channel Enable Buttons
Selectable Clock Reference
T
he Model 575 Digital Delay / Pulse Generator represents the latest in timing capabilities. With up to 8 outputs configurations as varied as the
applications the product serves, the Model 575 is clearly our most versatile instrument. We have combined advanced features such as a
Labview/USB interface, complex burst sequences, Divide-by-N, Setting Profiles, Dual Triggers, Dual Gating, Clock Divider, Pulse Picking and Negative
Delay with core technology in precision timing. Our 250pS Delay & Width resolution, and 200pS internal jitter, allow users great confidence in setting
up an experiment or synchronizing multiple events. Learn more about the newest Berkeley Nucleonics Pulse Generator…the Model 575.
Precision Instrumentation since 1963….
Flexible Gating Options
External Trigger Options
Individual Rates
The Model 575 is packed with gating options for almost any setup. You
may gate with a channel or on any input. You may gate individual
channels or gate all. Gate immediately (output inhibit) or gate after a
pulse (pulse inhibit).
Select channels for internal / external triggering, or free-run. Triggered
channels have flexible output choices: single pulse, burst at its clock
rate, continuous pulse train or a series of on/off pulses (duty cycle).
Each channel can have individual channel rates (either To or Tx... where
Tx is the alternate channel rate for that specific channel... e.g. T1 for
Channel 1). This is similar to having a separate clock for each output.
Multiplex Channels
Dual Triggers
Tek Stop
Trig’d
Channel D is the combination of Ch’s A, B &
C while not using its own width and delay
settings. However, Channel D is set up for
only TTL output
Two independant inputs may be used as
TriggerA and TriggerB. You can independantly
select a trigger for each channel.
BNC
MODEL
575
Range 0-5000s
Resolution 250ps
Timebase 25ppm (1ppm optional)
RMS Jitter 200ps
Pulse Inhibit Delay 150ns
Output Inhibit Delay 150ns
Number Up to 2
Rate DC to 1/(0.2us + longest delay)
Threshold 0.2 to 15 VDC
Maximum Input Voltage 60 V Peak
Resolution 10mV
Slope Rising or Falling
Impedance 1000 ohm
Jitter 800ps RMS
Insertion Delay 100ns
Number Up to 2
Threshold 0.2 to 15 VDC
Maximum Input Voltage 60 V Peak
Resolution 10mV
Polarity Active High/Active Low
Function Pulse Inhibit or Output Inhibit
Channel Behavior
Global w/ Individual Channel Enables
Rate 0.0002 Hz to 20.000 Mhz (40 MHZ optionl)
Resolution 5ns
Accuracy Same as timebase
Jitter 200ps
Settling 1 cycle
Burst Mode 1 to 10,000,000
Number 2, 4 or 8 Channel Outputs
Load 50ohm
Rise time (TTL) 3ns typ
Slew rate (Adjustable) 0.5 V/ns
Overshoot
<100mV + 10% of pulse amplitude
Levels TTL 0 to 4 VDC
*VAR adjustable amplitude,
2.0 to 20.0 VDC with 10mV
res, 20.0 VDC max transition
Number Up to 2
Rate DC to 1/(0.2us + longest delay)
Threshold 0.2 to 15 VDC
Maximum Input Voltage 60 V Peak
Resolution 10mV
Impedance 1Mohm + 40pF or 50ohm
Function(s)
Individual Channel Trigger/Gate/Follower
Trigger Slope Rising or Falling
Gate Polarity Active High or Active Low
Trigger Jitter < 2ns
Number 2, 4 or 8
Wavelength 820nm or 1310nm
Max Signal Rate 5 MBd
Max Link Distance 1.5km
Connector Type ST
Resolution 500ps
Accuracy 1 ns + .0001 x Delay
Number Up to 2
Wavelength 820nm or 1310nm
Max Signal Rate 5 Mbd
Max Link Distance 1.5km
Connector Type ST
Resolution 500ps
Accuracy 2 ns + .0001 x Delay
Optical Trigger 2412
Trigger Delay < 300ns
Jitter < 15ns
Communications USB/RS232
Global Gates/Triggers 2 Global Gate/Trigger Inputs
Channel Gates/Triggers
Optical/Electrial Available (5ns Jitter)
External Clock In 10 MHz - 100 MHZ
User Selectable in descrete values
External Clock Out 10 MHz - 100 Mhz
User Selectable in discrete values
Backwards Compatible
575-2C (2 Channel / RS-232 & USB)
575-4C (4 Channel / RS-232 & USB)
575-8C (8 Channel / RS-232 & USB)
575-2C-COM (2 Channel / Ethernet, GPIB, RS-232 & USB)
575-4C-COM (4 Channel / Ethernet, GPIB, RS-232 & USB)
575-8C-COM (8 Channel / Ethernet, GPIB, RS-232 & USB)
Contact Factory for High Voltage or Optical Options
Precision Instrumentation since 1963….
Delays