www.berkeleynucleonics.com 5.12 [SOURce]:SWEep Subsystem
[SOURce]:SWEep:DELay:AUTO
:SWEep:COUNt INFinite | <value>
[SOURce]:SWEep:COUNt Infinite | <value>
[SOURce]:SWEep:COUNt?
This command sets the number of sweep repetitions being played after triggering a sweep. If set to
INFinite, the sweep will be repeated until a [SOURce]:FREQuency:MODE or [SOURce]:POWer:MODE
command is issued.
*RST INFinite
Range INFinite or 2-65535
:SWEep:DIRection UP|DOWN|RANDom
[SOURce]:SWEep:DIRection UP | DOWN | RANDom
:SWEep:POINts
[SOURce]:SWEep:POINts <value>
[SOURce]:SWEep:POINts?
This command defines the number of step sweep points.
*RST 101
Range 2–65535
:SWEep:DWELl
[SOURce]:SWEep:DWELl <value>