Query the A coefficient.
[SENSe:]TEMPerature:SPRTD:BX {<value>|MINimum|MAXimum}
Set the B coefficient.
[SENSe:]TEMPerature:SPRTD:BX? [MINimum|MAXimum]
Query the B coefficient.
[SENSe:]TEMPerature:SPRTD:CX {<value>|MINimum|MAXimum}
Set the C coefficient.
[SENSe:]TEMPerature:SPRTD:CX? [MINimum|MAXimum]
Query the C coefficient.
[SENSe:]TEMPerature:SPRTD:DX {<value>|MINimum|MAXimum}
Set the D coefficient.
[SENSe:]TEMPerature:SPRTD:DX? [MINimum|MAXimum]
Query the D coefficient.
[SENSe:]TEMPerature:TRANsducer FRTD
Set the RTD Measurement to 4-Wire.
[SENSe:]TEMPerature:TRANsducer RTD
Set the RTD Measurement to 2-Wire.
[SENSe:]<function>:NPLCycles {0.02|0. 1|1|10|MINimum|MAXimum}
Set the integration time in number of power line cycles for the selected function.
This command is valid only for DCV, DCI, 2-wire ohms and 4-wire ohms.
[SENSe:]<function>:NPLCycles? [MINimum|MAXimum]
Query the integration time for the selected function.
[SENSe:]FREQuency:APERture {0.01|0.1|1|MINimum|MAXimum}
Set the gate time (or aperture time) for frequency function. Specify 10 ms (4.5
digits), 100 ms (default; 5.5 digits), or 1 second (6.5 digits).