Bnc 1201 - 6 1/2 Digit Digital Multimeter Manual de usuario Pagina 43

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 177
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 42
44
Sensor Type Temperature
Range(°C)
Temperature
Range(°F)
E -270~1000 -518~1832
J -210~1200 -140~2192
S -50~1768 -122~3236
T -270~400 -518~752
RTD (PT 100) -200~850 -392~1562
Table 3-1
3.7.1 Thermocouple Measurements
Connect the thermocouple adapter to the banana jacks on the front
panel. The difference between each type is subject to the patch thermal
leads.
Note: Only connection via the front panel can be used for temperature
measurements.
How to measure thermocouple
Use terminals switch to select terminals on the front panel.
To measure thermocouple must be via thermocouple adapter and the
low thermal patch leads as shown in Figure 3-12. (When the
mentioned modules are well connected as the following figure, please
go to the next step.)
Configure the thermocouple type and unit using CONFIG + SHIFT +
TEMP, such as K TYPE and ºC. When ready, press ENTER button.
Vista de pagina 42
1 2 ... 38 39 40 41 42 43 44 45 46 47 48 ... 176 177

Comentarios a estos manuales

Sin comentarios