receives :INIT[:IMM].
When :CALC2:LIM:CLE:AUTO is ON and whenever the counter
receives :INIT:CONT ON command, it :
o Invalidates the limit data,
o Clears the pass and fail counts (:CLAC2:LIM:FCO and :CALC2:LIM:PCO)
in the limit testing.
o Sets the Limit-Detect output to the in-limit voltage level.
When [:SENSe]:EVEN:LEV[:ABS]:AUTO is ON and :INIT[:IMM] command is
sent, the counter issues an auto-trigger on the source.
If the counter is in the process of making a measurement or
if :iNItiate:CONTinuous is set to ON, an :IMMediate command has no effect,
and error -231 (Init ignored) is generated.
Beginning a measurement or block of measurements with an :INIT[:IMM]
sets the pending Operation Flag to true. Completing the measurement or
block of measurements (normally or by aborting) sets Pending Operation
Flag to false.
For an automatically armed Totalize measurement, use :ABORt to terminate
the measurement initiated with :INIT[:IMM].
:INPut[1|2]:ATTenuation 1|10
Sets or queries the input attenuation.
Return
Numeric data transferred as ASCII bytes in <NR1> format.
Note
The input attenuation is set to X1 after *RST command.
:INPut[1|2]:COUPling AC|DC
Sets or queries the input coupling.
Return
A sequence of ASCII-encoded bytes: AC or DC.
Note
The input coupling is set to AC after *RST command.
:INPut[1|2]:FILTer[:LPASs][:STATe] <Boolean>