Internal 1 is used.
This command does not apply on Rise/Fall time, Duty cycle or Pulse width
measurements.
[:SENSe]:TINTerval:ARM:ESTARt[:LAYer[1]]:SOURce IMMediate TIMer
INTernal[1]
Sets or queries the start arm delay fro Time Interval measurements.
Return
A sequence of ASCII-encoded bytes. IMM means Immediate, TIM means Timer and
INT means Internal.
Note
This will be set to Immediate after the counter has been turned off or a remote
interface reset.
This command does not apply on Rise/Fall time, Pulse width or Duty Cycle
measurements.
[:SENSe]:TINTerval:ARM:ESTART[:Layer[1]]:TIMer <numeric value> [S]
Sets or queries the delay time in start arm for Time Interval measurements.
Return
Numeric data transferred as ASCII bytes in <NR1> format.
Notes
This will be set to 100E-9 seconds after the counter has been turned off or a remote
interface reset.
This command only applies when [:SENSe]:TINT:ARM:ESTART[:LAYer[1]]:SOUR
TIM is used.
This command dos not apply on Rise/Fall time, Pulse width or Duty cycle
measurements.
[:SENSe]:TINTerval:ARM:ESTOP:LAYer2:SLOPe POSitive NEGative
Sets or queries the slope of the external stop arm signal used in external arming Time
interval measurements.
Return
A sequence of ASCII-encoded bytes. POS means Positive edge and NEG means