<source list>
Default (@1)
Range (@1)|(@2)
:MEASure[:SCALar][:VOLTage]:MINimum? [(@1)|(@2)]
Measures voltage minimum.
<source list>
Default (@1)
Range (@1)|(@2)
:MEASure[:SCALar][:VOLTage]:NWIDth? [<reference>][,(@1)]
Sets the measurement arming to “auto” and measures negative pulse width.
The <reference> specifies the point on the pulse where the negative pulse width is determined.
This point can be specified in percentage or in voltage. The default unit is percent.
The <reference> parameter is used to configure the trigger setting. When it is specified
in percentage or default, auto-trigger is enabled and trigger level is the specified value.
When it is specified in voltage, auto-trigger is disabled and the trigger level is the
specified value.
<reference> default 50 pct
Percent range 0 to 100 pct
Percent resolution 10%
Voltage range -5.125V to 5.125V for ×1 attenuation
-51.25V to 51,25V for ×10 attenuation
Voltage resolution
0.0025V for ×1 attenuation
0.025V for ×10 attenuation
<source list> default (@1)
range (@1)
:MEASure[:SCALar][:VOLTage]:PWIDth? [<reference>][,(@1)]
Sets the measurement arming to “auto” and measures positive pulse width.
The <reference> specifies the point on the pulse where the positive pulse width is determined.
This point can be specified in percentage or in voltage. The default unit is percent.
The <reference> parameter is used to configure the trigger setting. When it is specified
in percentage or default, auto-trigger is enabled and trigger level is the specified value.
When it is specified in voltage, auto-trigger is disabled and the trigger level is the
specified value.
<reference> default 50 pct
Percent range 0 to 100 pct
Percent resolution 10%