Bnc 1105 Universal Counter Manual de usuario Pagina 33

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 197
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 32
33
3333
33
the trig Sens.
f. Use Show Status button to check the conditions of Input, Volt, Stats
and Limit.
g. Use Save&Recall button to save your settings (i.e., changes made in
steps a-d) if it’s needed (refer to Ch.4.9).
5. Press RUN button to make continuous measurements.
6. Press STOP/SINGLE button to stop current measurement or to make
single measurement.
Remote Interface Command
Use the following commands for making phase measurement:
[:SENSe]:FUNCtion[:ON] “[:][XNONe:]PHASe [1,2]”
Or
:CONFigure[:SCALar][:VOLTage]:PHASe [(@1),(@2)]
:READ[[:SCALar][:VOLTage]:PHASe]?
Or
:MEASure[:SCALar][:VOLTage]:PHASe? [(@1),(@2)]
3.10 Voltage Peak Measurement
The MODEL 1105 can also measure positive or negative voltage peak of
the input signal. Note that the Counter measures the input signal after
the signal is conditioned by the input settings (impedance, coupling,
attenuation, and filter, refer to Ch.4.7). The measurement result is
corrected for X10 attenuation. Also note that Gating&Arming, Limit and
Math are not available for voltage peak measurements.
To measure positive or negative voltage peak
1. Connect your signal source to channel 1 or 2.
2. Press VOLT PEAK button to select voltage peak measurement
function.
3. Press the softkey under VOLT to select input channel.
4. Optional settings:
a. Press the softkey under MODE to select the operational mode as
Fast or Slow.
b. Press the softkey under SHOW to select display of the result as
Vista de pagina 32
1 2 ... 28 29 30 31 32 33 34 35 36 37 38 ... 196 197

Comentarios a estos manuales

Sin comentarios