softkey under Save.). For more calibration information, please check
the List 4-1.
5. Press the softkey under Run Cal when everything is set ready.
Figure 4-27 Figure 4-28
Item
Softkeys Description
1.
Gain Inp1 + Run Cal
DC +5V Calibration at CH1
2.
Gain Inp2 + Run Cal
DC +5V Calibration at CH2
3.
Gain Att1 + Run Cal
DC +9V ATT x 10 Calibration at CH1
4.
Gain Att2 + Run Cal
DC +9V ATT x 10 Calibration at CH2
5.
Offs Inp1 + Run Cal
DC +0V Calibration at CH1
6.
Offs Inp2 + Run Cal
DC +0V Calibration at CH2
7.
Offs Att1 + Run Cal
DC +0V ATT x 10 Calibration at CH1
8.
Offs Att2 + Run Cal
DC +0V ATT x 10 Calibration at CH2
9.
T-F Quick + Run Cal
Time Domain Calibration (incl.
External Arm)
10.
T-F Fine1 + Run Cal
Time Interval Calibration
11.
T-F Fine2 + Run Cal
Time Interval Calibration ( incl.
External Arm)
12.
T-F Freq + Run Cal
10
MHz Calibration
13.
Save
Save the settings in the memory
List 4-1
Remote Interface Command
Use these commands to run calibration:
:CALibration:SECurity:STATe
<Boolean>,<NRf>
System Setup
Under the System Setup menu as shown in the Figure 4-29, users can
run self-test, adjust
the beeper setting, choose dot convention (i.e.,
using comma or period to separate the integral portion and fraction
portion of a number), and TOUT (i.e., If the time out “ON” is chosen, the
time out mode will be enabled, and as measurement time exceeds