Bnc 1105 Universal Counter Manual de usuario Pagina 95

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 197
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 94
95
9595
95
Width (150 mVpp with optional rear connectors)
>10 ns Pulse Width
50 mVpp to 10 Vpp
(150 mVpp with optional rear connectors)
Trigger Level(ATT x 1)
Range ±5.125 V
Accuracy ±(15 mV + 1% of trigger level)
Resolution 2.5mV
ATT x 10 Range X 10
Trigger Slope Positive or Negative
Auto Trigger Level
Range 0 to 100% in 1% steps
Frequency
Peak Voltage fast mode >10KHz
Peak Voltage slow mode > 100Hz
Amplitude> 100 mVpp (No amplitude modulation)
Damage Level
DC~400MHz 50 12 Vrms
0 to 3.5 kHz, 1 MΩ
350 V dc + ac pk
3.5 kHz to 100KHz,
1 MΩ
350 V dc + ac pk linearly derated to 12 Vrms
100KHz to 4
00MHz
1 MΩ
12 Vrms
Attenuator
Voltage Range x10
Trigger Range x10
1.
Specifications and Characteristics for Channels 1 and 2 are identical for both Common and Separate
Configurations.
2. Values shown are for x 1 attenuator setting.
Multiply all values by 10 (nominal) when using the x 10 attenuator
setting. Note that it may necessary to recalibrate the input offset in the application environment (especially at
high temperature) to achieve maximum sensitivity.
Channel 3 Input Specifications
Frequency Range Standard 375 MHz to 6 GHz
Channel 3 Input Characteristics
Impedance 50 Ω
Coupling AC
VSWR < 2.5:1
Power Range and Sensitivity (Sinusoid)
375 MHz to 500 MHz -16 dBm to +15 dBm
500 MHz to 1 GHz -20 dBm to +15 dBm
1 GHz to 2 GHz -23 dBm to +15 dBm
2 GHz to 4 GHz -25 dBm to +15 dBm
4 GHz to 5 GHz -21dBm to +15 dBm
5 GHz to 5.5 GHz -20 dBm to +15 dBm
5.5 GHz to 6 GHz -17 dBm to +15 dBm
Damage Level
Vista de pagina 94
1 2 ... 90 91 92 93 94 95 96 97 98 99 100 ... 196 197

Comentarios a estos manuales

Sin comentarios