A sequence of ASCII-encoded bytes. IMM means Immediate, INT means Internal2
and TIM means Timer.
Note
This will be set to Immediate after the counter has been turned off or a remote
interface reset.
This command does not apply on Rise/Fall time, Pulse width or Duty cycle
measurements.
[:SENSe]:TINTerval:ARM:ESTop[:Layer[1]]:TIMer <numeric value> [S]
Sets or queries the delay time used in stop arm for Time Interval measurements. The
range is 100E-9 to 10.000 seconds and the resolution is 100E-9 seconds.
Return
Numeric data transferred as ASCII bytes in <NR1> format.
Notes
This will be set to 100E-9 seconds after the counter has been turned off or a remote
interface reset.
This command only applies when [:SENSe]:TINT:ARM:ESTOP[:LAY[1]]:SOUR TIM is
used.
This command does not apply on Rise/Fall Time, Pulse width or Duty cycle
measurements.
Sets or queries
Gate
for Totalize measurements. For more information about
the front
panel gating settings and the interface command parameters, please refer to
the TOT table in Chapter 5.4.
[:SENSe]:TOTalize:ARM[:START]:SLOPe POSitive NEGative
Sets or queries the slope of the external start arm signal used in external arming Totalize
measurements.
Return
A sequence of ASCII-encoded bytes. POS means Positive and NEG means Negative.
Notes
This will be set to Positive after the counter has been turned off or a remote interface
reset.