Trigger Default setting is Trigger at 0 V
Results Range 1.5 ns to 10E+15
Resolution 1 count
Systematic
Uncertainty
± 1 count
Peak Volts
Results Range
-5.1 V to + 5.1 V
Resolution 2.5 mV
DC Signals 15 mV + 2 % of V
peak-to-peak amplitude greater than 200 mV
DC Signals (ATT x 10)
150 mV + 2 % of V
peak-to-peak amplitude greater than 1 V
1 Vp-p, 50 Ω, ATT OFF
100 Hz ~ 10 KHz
15 mV + 2 % of V
peak-to-peak amplitude greater than 200 mV
10 KHz ~ 5 MHz
15 mV + 4 % of V
5 MHz ~ 80 MHz
15 mV + 7 % of V
80 MHz ~ 300 MHz
15 mV + 15 % of V
Rear Input Option Channel Isolation
Frequency Front Channel(dB), typical Rear Channel(dB), typical
100KHz <-85 <-85
1MHz -85 -75
10MHz -78 -55
50MHz -67 -42
100MHz -62 -37
200MHz -55 -35
300MHz -50 -36
400MHz -47 -33
500MHz -44 -28
The peak volts measurement will keep operating up to 400 MHz, although results act as references only.
Tres is the resolution including effect of certain internal errors.
The differential channel Error terms which counted by many systematic uncertainty equations result
channel-to-channel disaccord and internal noise. These issues can be improved by
the TI calibration in the
well-controlled temperature environment.